UPDATES
18/09/2025
Two Blue-channel cameras exhibited vacuum leaks leading to elevated temperatures and potential condensation; the units were removed and sent for repair. For the remainder of 25B—and possibly early 26A—the Blue channels are offline in spectrographs 1A (edge of IFU field; MEF FITS ext. 3) and 4A (center of field; MEF FITS ext. 21). See attached map.
Further updates will be posted here.