MagE – Calibrations

MagE – Calibrations

Order definition 

 Fast 1×1 

Slit (“)0.50.70.851.01.21.52.05.0comments
Exp.Time (sec) 50353025201510 5orders 9-14 show counts between 15000-25000 (a couple of lines are saturated) – low counts in orders 6 and 18-20 (50 to 100 counts above background)
Exp.Time (sec)  99909060604030 15orders 18-20 have ~200-400 counts over the background – orders 10-14 show some saturation

Red Flats  

With lamp Qh – reaching 30000-40000 counts in orders 7-9 

Fast 1×1 

Slit (“)0.50.70.851.01.21.52.05.0
Exp.Time (sec) 50302520 181410 4

Slow 1×1

Slit (“)0.50.70.851.01.21.52.05.0
Exp.Time (sec) 5838 3225211713 5

Blue and Very blue flats : 

Lamp: Xe-flash 

Slit: 5″ 

Focus set to 1100  – remember to write down the nominal focus value, because it has to be manually inserted after flat sequence is over. 

Exp. time: 15s for Blue flats and 100s for Very blue flats 

A sequence of at least 15 blue flats and at least 5 Very blue flats is recommended. For more details take a look at the manual 