This manual describes how to take direct imaging CCD data on the Las Campanas 40-inch Swope telescope with the 4Kx4K E2V CCD231-84 detector. Last updated: October 19, 2026 (Nidia, William).
Activate telescope and instrument:
A set of switches and push-buttons has been installed on the upper left corner of the main desk:


- The telescope, the dome, the shutter, the windscreen, the dome lights and the tube fans are controlled from the Telescope window of the Swope application on swope-obs (see Telescope control). The Windows program TPS is only needed for a few things, listed in the TPS appendix.
- To activate the telescope, turn the ‘TCS & Console Enable’ key, wait 5 s, then set the ‘Console Power ON’ switch to ON. This powers the drives: the TCS initialises them by itself a few seconds later.
- Start the Swope application if it is not running: it is ‘Swope’ in the Dock of swope-obs. Its Telescope window opens with it. The ‘Link’ light in the Status box is green once the application talks to the TCS.
- Set the data path and the file number for the night: Camera → DataPaths, press ‘Default’ to fill in the night’s folders (utYYMMD1_D2, on the local disk and on DATA_SWOPE), then ‘Apply’; every frame is written to both. To change the number of the next frame, Camera → File Number unlocks the ‘File’ field of the camera window: type the number and press Enter. Existing files with that number are overwritten without warning.
- Turn on the dome lights: tick ‘Lights’ in the Dome and enclosure box of the Telescope window.
- Go up to the dome floor and check that nothing is in the way of the telescope or the dome: the platform down, no ladder, cable or tool near the telescope. The presets below move the telescope and turn the dome while nobody is watching them.
- Press ‘Cover…’ in the Target box and confirm. The telescope goes to the cover position and the dome to 180°.
- Go upstairs, raise the platform with the white hand paddle, and remove the telescope cover. The hand paddle power is enabled by the ON/OFF switch on the paddle, which has three functions, allowing movement of the platform, the dome and the wind screen. After the cover has been removed, move the platform all the way down and disable the hand paddle power. Go back to the control room downstairs.
- Press ‘Park…’ and confirm to bring the telescope back to Zenith.

- Untick ‘Lights’ to turn the dome lights off.
We recommend taking sky flats at twilight, which provide better correction (to 1% level) than domeflats. However, domeflats are useful for shutter and linearity tests. The ‘Flat…’ preset in the Target box of the Telescope window will set the telescope and the dome for domeflats. The quartz lamp is turned on from a box on top of the light-blue rack (do not exceed 935 counts).
Dome flats: exposure times and lamp settings by filter. The exposure times are a suggestion: check the level of the first flat and adapt them if needed.
| Filter | Exposure time | Lamp setting (counts) |
|---|---|---|
| V | 15 s | 900 |
| B | 80 s | 900 |
| g | 20 s | 900 |
| r | 10 s | 800 |
| i | 10 s | 700 |

Before opening the telescope:
- Turn on the fans: tick ‘Tube fans’ in the Dome and enclosure box.
- Lower the wind screen with ‘Down’ in the Windscreen row before opening the dome.
- Open the dome shutter with ‘Open…’ in the Shutter row (and close it with ‘Close…’). Preferably, do not open or close the dome with the telescope pointing to Zenith; ‘Open…’ warns when the telescope is above 80°.

- Tick ‘Auto’ (dome tracking) in the Dome and enclosure box. Telescope tracking does not need to be turned on by hand: the TCS starts it at the end of every slew (‘Send’ or ‘Slew…’). The ‘Tracking’ check box in the Status box shows it, and turns it off or on.
- If the ‘Remote’ light in the Status box is amber, the dome is in local control and ignores the dome, shutter and windscreen commands: press ‘Go Remote’ on the LPLC touch panel.
- The pointing model will be automatically loaded. If the TCS line of the Status box shows ‘Pointing model OFF’, turn it on in TPS or ask the staff.
- If the yellow ‘E-Stop Warning’ lamp is on, or a light of the Status box is red, see Troubleshooting.
The Dome and enclosure box of the Telescope window shows the dome azimuth, whether the shutter is open, the windscreen position and the dome lights, and the ‘Remote’ light in the Status box turns amber when the dome is in local control. The LPLC touch panel shows what the Telescope window cannot: whether the shutter is fully closed (the TCS reports only the open limit), each row of dome lights, and the dome PLC’s alarms (shutter or windscreen timeout, shutter overcurrent, dome stuck, barcode scanner, dome drive fault). Check it when a shutter, windscreen or dome command fails (see Troubleshooting). It is also where the dome is put back in Remote (‘Go Remote’).

Swope GUI
The Swope application on swope-obs consists of three main controls:
- Science Camera and Filter wheel control
- Guiders control
- Telescope control (telescope, dome and enclosure)
In addition to these GUIs there are the image viewer (Quick look tool), the dewar status window, the airmass display and the object list window. A password protect mode allows access to the engineering GUIs. A detailed description of the application can be found here: https://carnegie-observatories.github.io/swope/
The detector is read out through 4 amplifiers producing four independent files called ccdnnnnc1.fits – ccdnnnnc4.fits. The quadrants as seen in the Quick-Look tool are named counterclockwise beginning from the top right quarter of the image (i.e., top right is c1, top left is c2, bottom left is c3 and bottom right is c4).
Science Camera and Filter wheel control
This window controls the CCD camera and the filter wheel. Exposures are started with the Start button. Macros can be started with the Execute button.

Exposure parameters
ExpTime is the exposure time in seconds for the next observation. For this parameter and the ones for Loops and File# the entry window will be highlighted red after the number is typed. The value is not recorded by the program until the user hits Enter. The exposure time can be changed while the exposure runs, until 5 seconds before its end; the field is then locked until the readout is over. The new value applies at once, and a value shorter than the time already exposed ends the exposure. Remember to hit Enter to validate the new value.
Loops: Setting up a loop can be a convenient way to take flat fields and bias, or to make a series of observations in order to split the total exposure into shorter individual ones.
ExpType: The exposure type (saved in the fits header) can be set to “Object”, “Focus”, “Bias”, “Dark”, or “Flat”. Selecting “Bias” will automatically set the exposure time to 0 seconds. Selecting “Dark” will disable the shutter for the exposure. Selecting “Focus” will enable a focus sequence, automatically setting the loop number to 7 (this can be changed if desired). The focus frame will be saved under a normal filename (e.g. ccd0065.fits). In this frame, the spacing between the first and second images will be larger than for the rest of the sequence, allowing identification of the sequence order. Selecting “Focus (auto)” takes the same sequence with the telescope focus stepped automatically (see Focus).
Binning can be set to 1×1, 1×2, 2×1, 2×2 or 4×4. Note that the readout time given in this manual is for 1 x 1 binning.
Filter: There is room for 10 different filters in the filter wheel. Clicking on the button will present you with the available filters. A ‘combo’ option is also available to allow combination of two filters.
One or more Quadrants can be selected to be read instead of the whole detector (note that this will not save time because each quadrant is read independently). The selection is enabled by clicking the Q checkboxes
Object: Enter here the title you want for the image.
File: File number will be automatically incremented for every new exposure.
Execute: This command is used to start a macro. The macro file name must be specified on the next window to the right. This option is useful when shutter time and/or linearity tests are performed. More details here: https://carnegie-observatories.github.io/swope/Swope/scripts.html
Speed: only the Turbo readout speed is available at the present time.
Shutter: This is a flag that indicates whether the shutter is open or closed.
Snap will make a 4×4 binned image of the current field without incrementing the file number. The exposure time is as selected in ExpTime.
Subraster Allows to select subrasters on the chip. The regions can be defined giving its central coordinates followed by the width and height of the subraster (XYWH) or by its corners (XXYY). Savemode minimal will save only the selected region while savemode full will keep the original dimension of the quadrant filling with zero the region outside the selected.

Telescope control
The Telescope window commands the telescope, the dome and the enclosure, and shows the state of the TCS (the telescope control system), read about once a second. It opens with the application; if it is closed, Telescope → Telescope Window reopens it.

The boxes of the window:
- Time and Coordinates (left): UT, airmass and sidereal time, and the R.A. and Dec. (J2000) of the telescope. ‘Get’ reads them from the TCS; ‘Send’ slews to the coordinates typed in the fields, after checks and a confirmation.
- Offsets: the E/W and N/S steps in arcseconds, applied with the compass buttons. ‘Tonly’ moves the telescope only, ‘T+G’ also the guider probes. T+G has not been tested on the telescope with the guiders yet (tests TT23, TT24 and TT54 of the telescope test list): until it has, prefer ‘Tonly’ and find the guide stars again. ‘0,0’ undoes the accumulated offsets and ‘Zero’ resets their sums.
- Focus: the telescope (secondary mirror) focus, in encoder counts. Type a value and press Enter to move there, or press ‘+’ and ‘−’ to move by the typed amount. ‘Encoder’ shows the current position.
- Telescope sensors and Sky and weather: the tube and top-ring temperatures; the Sun and the Moon; the weather from the Swope station, with a ‘Weather’ light (advice only).
- Telescope (right): the position reported by the TCS: R.A., Dec., airmass, hour angle, altitude, azimuth, parallactic angle, LST and UT.
- Target: the TCS’s current target and its distance from the telescope; the presets ‘Park…’, ‘Flat…’ and ‘Cover…’; the CSET offsets, with ‘CSET…’ and ‘Reset…’.
- Dome and enclosure: the dome azimuth, the ‘In sync’ light and the ‘Auto’ check box (dome tracking); the shutter, with ‘Open…’, ‘Close…’ and ‘Stop’; the windscreen, with ‘Up’, ‘Down’ and ‘Stop’; the ‘Lights’ and ‘Tube fans’ check boxes; and ‘Go to az…’, which turns the dome to the azimuth typed next to it.
- Status: the lights: ‘Link’, the connection to the TCS; ‘Power’, the drives powered (amber: in stand-by); ‘Drives’, ready (amber while they initialise; red with a fault, a limit or no 24 V); ‘Remote’ (amber: the dome in local control); ‘Platform’ (red: not down); ‘Limits’ (red: a hard limit or an E-stop, named in its tooltip); ‘Fault’ (red: a hard fault of the TCS); ‘Slewing’ (amber while the telescope moves); ‘Ready’, whether a slew would be sent (red, with the reasons in its tooltip). Green: fine; grey: off; amber: a warning; red: a fault; hollow: not known. Then the ‘Tracking’ check box; ‘Gear’ (slow, fast or changing); ‘Age’, the time since the TCS last answered (amber when late, red when the link is lost). The ‘TCS’ line gives the TCS version and its clock against the Mac’s: ‘clock -0.1 s’ means the TCS clock is 0.1 s behind. The TCS computes where the sky is from that clock, so from 10 s off the line turns white on red (‘TCS clock off by … s’) and a notice asks to report it to the staff. ‘Pointing model OFF’ there means the TCS pointing model is off (see TPS).
- STOP (the red button, or ⌘.): stops a slew, a preset or an offset, tracking and the dome rotation at once, without asking. It does not stop the shutter or the windscreen, which have their own ‘Stop’ buttons. After a STOP, tick ‘Tracking’ and ‘Auto’ again.
How the controls behave:
- Buttons whose name ends in ‘…’ ask for confirmation before anything moves. The check boxes and the windscreen’s ‘Up’, ‘Down’ and ‘Stop’ act at once.
- A control is greyed while it cannot be used: the TCS would refuse it, a command of its kind is still running, or the TCS reading is not current. Hold the pointer over it, or over a light, to see why.
- Refusals and failures open a notice with the reason, and every event is listed in the Swope Messages window and written to the Swope log. What to do about the common ones is in Troubleshooting.
- Drive power is not in the application: the drives are powered from the switch box (‘TCS & Console Enable’ key, ‘Console Power ON’ switch), and the TCS initialises them by itself. The ‘Power’, ‘Drives’ and ‘Gear’ lights in the Status box show it. If ‘Power’ is amber (‘drives in stand-by’), the drives were put in stand-by in TPS and must be switched on there again.
Every control is described in the application documentation.
Object List
The Object List window (Telescope → ObjectList) holds the targets of an object list file, opened with File → Open ObjectList; the pointing stars are in ‘cset.objects’. Select an object and press ‘Slew…’ to go there, with the same checks and confirmation as ‘Send’; after the slew, the object’s name and comment are copied into the camera’s Object and Comment fields. ‘Preview’ only copies the object’s coordinates into the Coordinates box of the Telescope window, where ‘Send’ slews there. ‘Add’, ‘Edit’ and ‘Remove’ change the list, and File → Save As… writes it to disk. The file format is given in the appendix.

Pointing
- Open the list of pointing stars in the Swope application (File → Open ObjectList, ‘cset.objects’) and pick a star close to Zenith. Select it and press ‘Slew…’ in the Object List window.
- A dialog shows the target’s position, altitude, airmass and hour angle, how long it can be tracked, the Moon distance, how far the telescope has to go, and any warnings; press ‘Slew’. Alternatively, type the R.A. and Dec. (e.g. 13:52:00.9 and -31:37:10.0, equinox 2000) in the Coordinates box of the Telescope window and press ‘Send’, which asks the same way.

- The ‘Slewing’ light in the Status box is amber while the telescope moves, and the Mac says ‘target reached’ when it arrives. If the slew is refused or fails, a notice gives the reason.
- To look up a star by name instead (Simbad), use TPS; the Target box of the Telescope window shows the target set there.
- Take a short exposure (1-2 s) of the field and compute the required offsets to place the bright star at the center of the image (scale is 0.435 “/pixel). Use the Offsets box of the Telescope window to apply these offsets. Take a new exposure and repeat procedure until pointing is right.
- Press ‘CSET…’ in the Target box of the Telescope window and confirm; the dialog shows the CSET HA and Dec. before and after. The telescope must be tracking and within 1° of the target. The 1° is a check of the Swope application, not of the TCS, which only needs a target to be set: a CSET of more than 1° would also move the TCS’s limits by that much. If the star really needed more than 1° of offsets, do not CSET; tell the staff (engineering mode has ‘Modify CSET…’). ‘Reset…’ sets both CSET values back to zero.

- Next click on Centerfield on the guider window and check that the bright star ends near the center of the guider display. If it doesn’t, move the guider to place it near the center and then click on SetCenter on the guider window; a dialog-box will ask you for confirmation. Do not forget to click ‘Park’ on the guider window after this, to take the guider probe away from the Centerfield.
Focus
The current focus value is shown in the Encoder field of the Focus box in the Telescope window. That box works in encoder counts: a focus of 23.125 in the units used below is 23125 there, and a step of 0.050 is 50.
Focus sequences are allowed by setting Focus as ExpType on the CCD command window. This will allow 7 position sequences (Loops =7). This default can be changed to other number if desired. We recommend taking 10s exposures changing focus by 0.050 units in between. The secondary mirror is moved from the Focus section of the Telescope window (see Telescope control). The focus needs to be determined for each filter to be used; a focus value of 23.125 was found for V filter after November 2023 engineering (for that setting, the guider focus was 3700). Focus offsets for the different filters are displayed in the board (a value -0.450 relative to V is good for all the Sloan filters). Once the focus sequence is over the corresponding file can be displayed and examined in the IRAF window. The spacing between focus exposures will be double between the first and second steps to allow identification of the sequence order. Once the approximate focus value is known, it is possible to run new focus sequences using smaller focus steps (such as 0.025 or 0.020). Focus may change along the night depending on temperature and telescope position. Sometimes a new focus sequence will be needed.
Focus offsets by filter, relative to V, in encoder counts (100 counts are 0.100 in the units above). V, B and the Sloan values are the ones the focus predictor script prints; the others come from the 2026 focus sweeps and from the nightly reports. Check the focus after a filter change, and after a change of filter set: other B filters used in the past had a different offset (in 2022 an observer used B −100 and Sloan −550, before the current B was installed).
| Filter | Offset from V (counts) |
|---|---|
| V (9844, Johnson) | 0 (the reference) |
| B (LC3013) | +100 |
| Sloan u’, g’, r’, i’ | −450 (r’ measured at −461 ± 9 in 2026) |
| Sloan z’ | −450 assumed, as the other Sloan filters; not measured |
| U, B, V, R, I of the parfocal set (1525 to 1529, installed December 2023) | about 0: parfocal with each other and within the nightly drift of V |
| Strömgren y | +33 (2026 focus sweeps) |
| R (LC3015) | +47 (2026 focus sweeps) |
| Hα 657 | +64 (2026 focus sweeps) |
| WN | −110 (2026, few sweeps) |
| Washington C | −281 (2026, few sweeps) |
| Narrow-band 6780 and 6830 | 0 reported in 2022; +67 and −130 from very few 2026 sweeps |
If you are using several filters, and already know the focus offsets between them, you may edit the Filter Hardhat (available under the Camera menu) to set the corresponding offsets so the guider focus follows automatically each time you switch filters (the telescope focus itself is set in the Focus box of the Telescope window). The filter Hardhat also provides a ‘Combo’ option which allows superposition of two different filters (one from each filter wheel).

Focus (auto): with “Focus (auto)” as ExpType, the application steps the telescope focus itself. ‘Start’ asks for the centre and the step of the sweep, in encoder counts; the centre defaults to the current focus and the step to the last one used (50 at first). The sub-exposures are taken from the lowest focus upwards: the first one is (Loops − 1)/2 steps below the centre, rounded down, and each following one a step higher. The frame is the same charge-shuffled focus frame as with “Focus”.

Focus predictor script by Chris Burns:
There is a script provided by Chris Burns that allows predicting the right focus based on air mass and temperature. The script is in /Users/obs1/Scripts on swope-obs.
To run it, type:
%~/Scripts/SwopeFocus [-h] [-plot] [-pars] temp AM
Where temp is the temperature displayed in the Swope weather station and AM is the air mass. The -plot and -pars options will issue a plot of the fit and report the fitting parameters.
The current expression to compute focus for the V filter is:
Focus (V) = 23.193 + 0.0291 * (T – 15) + 0.0936 * (AM – 1) + 0.00024 * (JD – 2461000)
where JD is the Julian Date, which the script takes from the computer clock.
Guiding
The main components of the guiding interface are the Skymap, Guider1 and Guider2 windows.
Skymap displays a star catalog and an overlay of the science camera field of view (red box) the patrol area of the guiders (blue boxes) and the area where the guiders vignette the science array (dashed rectangle). By selecting a star the catalog entry will be displayed in the text box in the lower left of the window. The selection will also update the probe relative position on the Guider window. The probe will be moved to that position by pressing the Goto button. The guider probe can be also moved by selecting an offset or by using the keypad (its speed, from Slow to Turbo, is chosen in the box next to it; Disabled turns the keypad off). The last section of the Guider window allows to select imaging mode (i.e. the guiding mode) or Shack-Hartmann (not yet available). The focus of the guiding camera can be changed using the text box in the last row or by moving the slider.
The star catalog in use by Skymap is indicated in the top left box; which can be clicked to see the available options and eventually switch catalogs.
The SkyMap and the airmass display follow the telescope position known to the Swope application. It is updated by ‘Get’, when an exposure (also a snap) starts, and after a slew, a preset or an offset made from the application. After a slew made from TPS, press ‘Get’ or take an image.

Once the probe is at the desired position it should be possible to see the guiding star appear on one of the guider displays (see figure).
The guider focus can be adjusted by clicking and moving the slide bar button on the guider GUI or just writing the desired focus value next to it. Each time the telescope focus is modified (e.g. to apply focus offsets between filters) the guider focus has to be modified accordingly.
The keyboard below the guider display is used to interact with the guider computer.

At the beginning of the night a software reset of the camera is recommended by typing DSPI. It is also recommended to check that the PA and PX parameters are set according to the values in the Guider table of the Appendix. The cooling system is initiated by entering TEC 1.5. To adjust the range of intensities between black and white displayed the command SPAN is used (e.g. SPAN 1000). In order to guide, the guiding box has to be moved on the star image using the cursors. To change cursor steps use the keys ‘help’ and ‘delete’.
Guiding is started with <F3> and it is stopped with <F1>. Usual guider settings are ‘sn’ = 0.5, ‘bx’ between 19 and 23, ‘tf’ = 2, ‘tg’ = 0.60, ‘avf’ = 0, ‘avg’ = 1 or 2. Span around 1000 will work for a 10th mag. guide star; depending on seeing and transparency.
A more detailed description of the guider software (GCAM) can be found at this address: https://www.lco.cl/wp-content/uploads/telescopes/magellan/gcam-usermanual.pdf

Check lists
Beginning of the night
- Turn on telescope/instrument power and TCS (rotate the ‘TCS & Console Enable’ key).
- Wait 5 s, then set the ‘Console Power ON’ switch to ON.
- Start the Swope application (‘Swope’ in the Dock of swope-obs) if it is not running, and check that the ‘Link’ light of the Telescope window is green.
- Move the telescope to the Cover position (‘Cover…’ in the Telescope window) and remove the telescope cover.
- Lower wind screen (‘Down’ in the Windscreen row).
- Open dome shutter (‘Open…’ in the Shutter row).
- Turn on the fans (tick ‘Tube fans’).
- Make sure that dome lights are off (‘Lights’ unticked).
- Tick ‘Auto’ (dome tracking, Dome and enclosure box). Telescope tracking starts by itself when you go to a target.
- Turn on guider cooling (TEC 1.5).
End of night
- Make sure the guider is ‘off’ (F1) and set it to the Park position.
- On the guider console type Tec 0.0 to stop cooling and SPAN 9999999 (or any large number).
- Untick ‘Tracking’ (and ‘Auto’, for dome tracking, if you wish) in the Telescope window.
- Close the dome shutter (‘Close…’) and check on the LPLC touch panel that it reads closed; then set the wind screen up (‘Up’).
- Move the telescope to the Cover position (‘Cover…’).
- Turn the dome lights on (tick ‘Lights’).
- Go upstairs and put the telescope cover on. Take the platform all the way down.
- Turn off the dome lights (untick ‘Lights’).
- Press ‘Park…’ to move the telescope to Zenith.
- ‘Park…’ also turns the dome to 180 degrees. If ‘Dome az.’ does not read 180, type 180 in the field next to ‘Go to az…’ and press ‘Go to az…’.
- Dome tracking is off after ‘Park…’: check that ‘Auto’ is unticked.
- Turn the fans off (untick ‘Tube fans’).
- Set the ‘Console Power ON’ switch to OFF.
- Turn off the telescope and TCS by rotating the ‘TCS & Console Enable’ key.
- Activate screen saver in the LPLC panel.
- Fill out the C40 daily report at https://reports.lco.cl (VPN required).
- Before leaving the telescope please check that no bird is trapped within the dome.
Troubleshooting
A command of the Telescope window that is refused or fails opens a notice with the reason, and the Swope Messages window lists every event. These are the common problems, by subsystem, and what to do. If in doubt, call the technicians.
Power, E-stop and drives
- The yellow ‘E-Stop Warning’ lamp is on (after an E-stop or a power failure): release the E-stop button if it was pressed, then press the black ‘E-Stop Reset’ button to restore power to the telescope and the instrument. Until then the ‘Drives’ light of the Telescope window is red (‘no 24 V: key switch or E-stop’), and the ‘Limits’ light red when the dome or control-room E-stop is reported.
- ‘Drives’ red with ‘no 24 V’, and no E-stop: turn the ‘TCS & Console Enable’ key on, wait 5 s, then set the ‘Console Power ON’ switch to ON.
- Switching the key off and on: wait at least 15 s between off and on. A faster cycle trips the dome drive, and the dome then does not turn until the staff reset it.
- ‘Power’ amber (‘drives in stand-by’): the drives were put in stand-by in TPS; ask the staff to switch them on there (Settings menu, ‘Telescope Power’).
- ‘Fault’ red (a hard fault): after a power cut, or when the TCS program was started twice, only a restart of the TCS clears it (see Restarting the TCS).
Telescope
- ‘Link’ red, notice ‘TCS link lost’: the application keeps trying, and the Swope Messages window says ‘TCS link back’ when it answers again; meanwhile the window shows the last values in grey. If the link does not come back within a minute or two, restart the TCS (below).
- A slew is refused because the telescope is below 15°, past the pole or beyond ±7 h: bring the telescope back with the hand paddle.
- ‘the telescope did not start moving’, or ‘stopped … from the target’: press ‘Get’, then ‘Send’ again; another program may have changed the target.
- ‘Limits’ red: its tooltip names the limit. An inclination, ±HA or ±Dec limit: the staff recover the telescope. ‘HA holder not engaged’ or ‘Dec holder not engaged’: engage that axis’s handle. An E-stop: see above.
- ‘Pointing model OFF’ in the TCS line: the telescope points less accurately; turn the model on in TPS (Pointing menu, ‘Tpoint model’), or ask the staff.
- ‘TCS clock off by … s’: the clock of the telescope computer is wrong, so the pointing and the limits are wrong too; report it to the staff.
Restarting the TCS
The TCS runs on its own computer. Its output can be watched, and the program restarted, from the Windows PC:
- On the Windows PC, open Bitvise SSH Client. It will automatically load the SSH profile for the TCS (IP: 10.8.80.53, user/password as usual for Swope PC).
- Click Login, then open a New Terminal Console.
- In the new terminal console, type: tmux attach, and then enter.
- This displays the TCS output, including guider and telescope communication messages: for example, Guider Xnn Ynn Gnnnn (where ‘n’ is a number) indicates that the guider and telescope are communicating.
- To stop the TCS, press CTRL + C. To restart it: cd TCS, then ./SwopeTCS.
- Closing the terminal does not affect the TCS.
- If in doubt, contact the technicians.
Dome and enclosure
- ‘Remote’ amber: the dome is in local control and ignores the dome, shutter and windscreen commands (‘Auto’ can still be unticked, and the Stop buttons and ‘Lights’ still work). Press ‘Go Remote’ on the LPLC touch panel.
- A shutter, windscreen or dome command fails: look at the LPLC touch panel, which shows the dome’s own alarms (timeout, overcurrent, stuck, drive fault).
- ‘shutter open failed’ or ‘shutter close failed’: the dome stopped the shutter (after 30 s, or because its motor drew too much current), and the application has sent ‘Stop’. Call the staff, who can close the shutter from the LPLC panel; do not keep trying.
- ‘windscreen up failed’ (or down): ‘did not move’: try once more. ‘stopped at … %’: it is stuck, or hit its timeout; look at the LPLC panel and call the staff.
- ‘dome to … failed, did not move’: the dome drive has tripped, for example after the console key was switched off and on in less than 15 s; the staff reset it. ‘stopped at …°’: press ‘Go to az…’ again, and if it repeats, call the staff.
- ‘In sync’ stays amber with ‘Auto’ ticked: the dome stopped short and is not turned again. Type the ‘req.’ value in the field next to ‘Go to az…’, press ‘Go to az…’, then tick ‘Auto’.
The dome does not respond at all: if the dome becomes fully non-responsive (which happens very seldom), it may need a reset from the dome control box located in the equipment room:

Camera
- ‘DSP error’, ‘line errors from DSP’ or ‘no data from ccdserver’: try again; if it persists, open Camera → CCD Hardhat and press ‘Reset DSP/CCD-Server’.
- ‘ccdserver: too late’: the exposure is within 5 s of its end; it can no longer be paused or have its time changed.
Computers and remote display
The Windows remote display looks like the picture below: double click on it and a dialogue will open indicating that the system has been disconnected and it will reconnect immediately.

Appendix
Tables
| Quantum Efficiency | |
|---|---|
| 350 nm | 42.1 % |
| 400 nm | 79.7 % |
| 500 nm | 85.8 % |
| 650 nm | 92.0 % |
| 900 nm | 62.5 % |
| Quadrant | Gain (sigma) | Readout Noise (sigma) |
|---|---|---|
| CCD1 | 1.045 (0.029) | 3.026 (0.093) |
| CCD2 | 1.018 (0.028) | 3.206 (0.102) |
| CCD3 | 1.003 (0.027) | 3.414 (0.103) |
| CCD4 | 1.029 (0.029) | 3.154 (0.099) |
| Detector | E2V CCD231-84 |
|---|---|
| Pixel scale | 0.435 arcsec/pixel |
| Readout time | 37 s |
| Field of view | 29.7 X 29.8 arcminutes |
| Guider # | PA | PX | FOV |
|---|---|---|---|
| 1 | -210 | 0.163 | 2.78 arcmin |
| 2 | -30 | 0.163 | 2.78 arcmin |
Pointing limits
Current pointing limits at Swope are:
Altitude: 20 degrees (airmass 2.9).
Hour Angle East: -6 h
Hour Angle West: +7 h
The Telescope window checks every command before it is sent. When a limit is not met the command is refused, with a notice that says why; the slew dialog warns about the rest. The 20° above is the observing limit: the window warns below it and refuses below 15°.
| Limit | In the Telescope window |
|---|---|
| Target altitude | refused below 15° (airmass 3.8); a warning in the slew dialog below 20° |
| Target hour angle | refused outside −6 h … +7 h; the dialog gives the time left until +7 h or 15° |
| Target declination | refused below −89° |
| Telescope position | a slew is refused while the telescope is below 15°, past the pole or beyond ±7 h: bring it back with the hand paddle. Tracking stops by itself below 15° or past +7 h |
| Slew | refused without drive power, in stand-by, with a hard fault, a hard limit or an E-stop, with the platform up, or while the gear changes; a slew not finished in 10 minutes is stopped |
| Offsets | only while tracking in slow gear, one at a time; up to 3600″ in Dec. and 1800″ of hour angle in R.A.; T+G offsets up to 600″ |
| CSET | within 1° of the target; ‘Modify CSET…’ changes it by at most 1° |
| Focus | 20000 to 28000 encoder counts (not enforced in engineering mode) |
| Shutter and windscreen | refused while the dome is in local control; ‘Open…’ warns above 80° and with a red weather advisory; a shutter that has not reached its end after 30 s, or a windscreen after about 67 s, is stopped by the dome’s own safety |
| Hard limits | a limit switch or an E-stop cuts the drives: the ‘Limits’ and ‘Drives’ lights show it, and every move is refused |
Object lists
Object lists for the Swope application are text files, opened with File → Open ObjectList. Their default folder on swope-obs is /Users/Shared/Library/Application Support/Swope/Objects/. Each line holds one object, with its fields separated by ‘|’: name, R.A., Dec., equinox and a comment.
Example (file ‘targets.objects’):
JGEM24a | 08:08:51.8 | -17:30:59.3 | 2000.0 |
AT2024hfx | 08:19:41.6 | -17:28:07.5 | 2000.0 |
AT2024hfq | 08:25:08.9 | -24:43:45.9 | 2000.0 |
Select an object and press ‘Slew…’ to go there, or ‘Preview’ to copy its coordinates into the Coordinates box of the Telescope window (then ‘Send’). Objects can be added, edited and removed in the Object List window, and the list saved with File → Save As….
TPS reads its own object lists, in a different format: see TPS.
Macros
Macros can be used by placing a text file in the DATA directory of the observer’s account on swope-obs.
The file could be something like this:
filter g
exptime 12
start
filter B
exptime 30
start
filter u
exptime 100
start
To run the macro, the file name has to be specified in the window (usually empty) next to Execute on the CCD GUI, for example: /Users/obs1/DATA/macro_test Then the macro is started by clicking on the Execute button.
TPS (the Windows telescope program)
TPS is the telescope program of the TCS author. It runs on a Windows machine whose remote display is shown on the right side monitor of swope-obs, and it talks to the same TCS as the Swope application: both can be connected at once, and a target or a command given in one shows in the other. The Telescope window of the Swope application now does everything needed to observe (see Telescope control); TPS is still used for:
- looking up a target by name (Simbad), see below;
- turning the pointing model on, when the Status box of the Telescope window shows ‘Pointing model OFF’ (Pointing menu, ‘Tpoint model’);
- putting the drives in stand-by and back (Settings menu, ‘Telescope Power’). This is not needed to observe: the drives are powered from the switch box (see Activate telescope and instrument). Only if the ‘Power’ light of the Telescope window is amber (‘drives in stand-by’) must they be switched on here;
- as a fallback for the Telescope window: the KeyPad (‘Go to Target’, ‘STOP Dome Telescope’, ‘Tracking’, ‘Dome Tracking’, ‘Light’, ‘Tube Fans’) and the Dome menu (‘Open’, ‘Close’, ‘STOP’, ‘Dome Rotation’, ‘Wind Screen Up/Down/Stop’) do the same as the window’s controls.
The Windows machine is reached from swope-obs with the Windows App (Microsoft’s remote desktop client), which is in the Dock: open it and double-click the one PC it lists; the user and password are the usual ones for the Swope PC. Its desktop opens on the right side monitor. If it shows the ‘disconnected’ picture (see Troubleshooting), double-click it and it reconnects.
If the TPS window is not open, double click on the TPS icon in the Windows display. This will bring up the TCS GUI and the KeyPad window (see figures below). The TCS GUI has several menus; ‘Pointing’ and ‘Dome’ are the most used.


Looking up a target by name
- On the TCS Pointing menu click on ‘Select object’ to open the ‘Select the telescope target’ window:

- Click on the bullet next to ‘Simbad’ and write the star’s name in the window next to it (e.g. HR5212, or hr5212), then click on Search. The object’s details will be displayed. Click ‘OK’.
- Alternatively, you can click the bullet next to ‘Coordinates’ and manually enter RA and DEC (e.g. 13 52 00.90 -31 37 10.0).
- In the KeyPad, click on ‘Go to target’. The message ‘Telescope slewing’ will appear in red in the TCS GUI. This message will disappear once the target position is reached. The Target box of the Telescope window shows the same target, and the offsets, ‘CSET…’ and the rest of the window work as usual once the telescope is there.
TPS object lists
Object lists can be loaded in the TPS directory of the Windows machine (use a pendrive or connect from swope-obs via SAMBA -ask the technicians).
At present object lists need to be in the following format (this may be changed in the future):
%Object_name%RA(degrees and fraction)%DEC(degrees and fraction), no spaces.
Example:
Filename ‘targets.dat’
%JGEM24a%122.215846%-17.516465
%AT2024hfx%124.923208%-17.4687389
%AT2024hfq%126.286902%-24.729418
%AT2024hfs%127.160212%-20.383216
%AT2024hfr%130.266165%-18.593131
In the TPS directory of the Windows machine, edit the file List.dat and add as a new line the filename of the object list you want to load. Both List.dat and targets.dat are in the TPS directory.
Then close TPS and open it again so the new list can be read.
In the Pointing menu, go to ‘Select target’ and click on the bullet next to ‘Custom object’. The available target lists will be displayed (List.dat file). Choose the list you want to load and click on the next window to select the target. The target details will be displayed. Click OK and then Go to Target.