LDSS3 – Calibrations

How Can We Help?

LDSS3 – Calibrations

Here we provide some reference numbers for calibrations with LDSS3-C

Spectroscopy 

Below we provide some reference numbers for exposure time and suggested quartz lamps in order to obtain a flux level between 30000 and 35000 ADU, on amplifier1 (C1) , with various longslit widths for the standard setup of: speed:fast / gain:low / binning:1x1

Grism(lamp) / Slits 0.75 - center1.0 - center1.25 - center1.5 - center
 VPH - All (Qh)   11 sec   9 sec    6 sec    6 sec
 VPH - Red (Qh)  10 sec   8 sec    6 sec    6 sec
 VPH - Blue (Qh)  85 sec  70 sec  50 sec  50 sec

Imaging 

In the table below we provide some reference numbers of imaging flat counts obtained in the g'r'i'z' filters and suggested lamp intensities (updated Jan2023):

FilterCounts (Fast bin1x1)Exposure time (sec) Voltage (V)
  B        22000          10     3.5 + Ql
  g'        31000          15     2.5
  r'        34000          10     1.7
  i'        32000          10     1.2
  z'        33000          9     1.0