IMACS – Calibrations

Numbers below correspond to new configuration - after the dewar swap 

Since the end of June, we are experiences issues with the power supply of the Variable Lamp, thus the tabulated voltages might not give the expected level of counts. To get sufficient counts for the various configurations it is suggested that you use a voltage in the range of 3.0-4.7V.

Imaging

In the table below we provide some reference numbers of imaging flat counts obtained in the various filters and suggested lamp intensities (Ql and Qh refer to the Quartz lamps in the DCU). These frames were obtained with the dome closed at night. 


Note: During daytime the dome is not completely light tight and the exposure times should be kept <10sec in order to minimise fluctuations due to stray light:


Camera f/2 :

FilterCounts (Fast bin1x1)Exposure time (sec) Voltage (V)
  Bessel - B        28000          6     Ql
  Bessel - V        27000          8     3.0
  Bessel - R        28000          3     2.5
  CTIO - I         30000          5     2.0
  sloan - u
  (new)
        20000          55     Qh+Ql
  sloan - g        25000          7     3.5
  sloan - r        26000          5     2.5
  sloan - i        27000          5     2.0
  sloan - z         26000          3     2.0
WB 4800-7800 30000 5 2.2


Camera f/4 :

FilterCounts (Fast bin1x1)Exposure time (sec) Voltage (V)
  Bessel - B        28000          2     Qh
  Bessel - V        30000          4     4.0
  Bessel - R        24000          2     3.0
  CTIO - I         25000          2     2.5
  sloan - u
  (new)
        15000          70     Qh+Ql
  sloan - g        29000          3     Ql 
  sloan - r        30000          4     3.0
  sloan - i        30000          3     2.5
  sloan - z         26000          2     2.5

 
Flats of the sloan-u are not possible with current lamps (120sec with Ql+Qh give ~4500counts with f2 and ~2500counts with f4). For this filter twilight flats should be obtained. 

A new sloan-u filter has been purchased that is more efficient than the previous one. Dome flat calibrations are possible using available lamps.

Spectroscopy 

Camera f/2 :

The following exposure times will give between 20000-25000 counts using the Qh lamp in the DCU and the Spectroscopic2 filter 

Grism/Slit0.7"0.9"1.2"1.5"2.5"
 150_10.8 5sec 4sec  3sec 2sec 1sec
 200_15.0 7sec 5sec 4sec 3sec 2sec
 300_17.5 15sec11sec 9sec 7sec 4sec
 300_26.7 11sec 8sec 6sec 5sec 3sec

Camera f/4 :

The following exposure times will give between 20000-22000 counts using the Qh + Ql lamps in the DCU 

Grating (tilt) /Slit0.7"0.9"1.2"1.5"2.5"
 150-3.4 (3.4)  11sec 9sec 7sec 5sec 3sec
 300-4.3 (6.0)  40sec 30sec 20sec 16sec 12sec
 600-6.54(8.0)  sec sec sec sec sec
 600-8.6 (9.3) 80sec 60sec 50sec 40sec 25sec
 600-13.0 (14.0) 50sec 35sec 25sec 20sec 15sec
 1200-17.5 (19.0)  300sec 250sec 150sec 125sec 90sec
 1200-26.7 (27.32)  100sec 85sec 60sec 50sec 35sec
 1200-26.7 (33.5) 120sec 100sec 70sec 55sec 40sec
 1200-32.0 (32.2)  sec sec sec sec sec